首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >Effects of thermal regulation structures on the photon sensitivity and spatial resolution of a 1 mm3 resolution breast-dedicated PET system
【24h】

Effects of thermal regulation structures on the photon sensitivity and spatial resolution of a 1 mm3 resolution breast-dedicated PET system

机译:热调节结构对1 mm 3 分辨率宠物专用PET系统的光子敏感性和空间分辨率的影响

获取原文

摘要

This work presents the effect of special mechanical structures inter-dispersed between detectors on the photon sensitivity and spatial resolution of a 1 mm3 resolution breast dedicated PET scanner. A degradation of (7–8)% on the single photon sensitivity and 15% on the coincidence photon sensitivity has been estimated from simulated data. However, no significant effect on both intrinsic and reconstructed spatial resolution was predicted by simulations. A reduction on the number of the multi-interaction photon events has also been observed which leads to a minor improvement on spatial resolution under the presence of additional material surrounding the LYSO-PSAPD detectors. Even though the analysis presented above used a rather inaccurate positioning scheme of multi-interaction photon events among detectors, more sophisticated algorithms are currently under development in order to properly identify and include these events in the analysis [5]. Currently Al structures that will assure a good compromise between mechanical robustness, temperature stability and satisfactory system performance are under construction
机译:这项工作介绍了特殊机械结构在探测器之间分散的特殊机械结构对1mm3分辨率乳房专用PET扫描仪的光子敏感性和空间分辨率之间的分散。已经从模拟数据估计了单光子敏感度对单光子敏感性的(7-8)%的降解和15%的偏差。然而,通过模拟预测了对固有和重建的空间分辨率的显着影响。还观察到多相相互作用的数量的减少,这导致在围绕Lyso-PSAPD检测器的附加材料的存在下对空间分辨率进行微小的改进。尽管上面呈现的分析使用了检测器之间的多交互光子事件的相当不准确的定位方案,但是更复杂的算法目前正在开发中,以便适当地识别和包括这些事件[5]中的这些事件。目前,可以确保机械稳健性,温度稳定性和令人满意的系统性能之间的良好折衷

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号