首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >Evaluation of the HD and HD+TOF reconstructions for Siemens' Biograph-mCT TOF PET scanner
【24h】

Evaluation of the HD and HD+TOF reconstructions for Siemens' Biograph-mCT TOF PET scanner

机译:评估西门子传记 - MCT TOF PET扫描仪的HD和HD + TOF重建

获取原文

摘要

We describe an evaluation of the HD and HD+TOF reconstructions in terms of contrast vs noise, precision vs accuracy, and inter-realization variation on Siemens' Biograph-mCT time-of-flight (TOF) PET scanner. The HD-PET reconstruction is the 3D-OSEM with point-spread-function (PSF) modeling, and the HD+TOF is the HD-PET with the time-spread-function (TSF) modeling based on the TOF information acquired from the scanner. A chest-spine phantom was scanned, and the data were reconstructed using the HD and HD+TOF methods. Hot contrast as well as cold contrast from regions with different densities was evaluated as a function of the voxel-noise in the images. The coefficient of variation (COV) in contrast was evaluated over 10 realizations at low statistics and plotted as a function of the bias in contrast. The standard deviation (STD) images were also compared between the HD and HD+TOF reconstructions. It was observed that the HD+TOF reconstruction outperforms the HD method in terms of hot contrast vs noise as well as cold contrast in the low density regions. However, for the high density regions the HD+TOF showed slightly worse results as compared to the HD method. Consistent results were also observed in terms of precision vs accuracy. The STD image comparison showed similar results between the HD and HD+TOF reconstructions. Additionally, the phantom data were also reconstructed with various methods and parameters for both Biograph-mCT (mCT) and Biograph-40 (B40), and the contrast vs noise evaluation showed that the same reconstruction method with the same or similar parameters on both scanners performs similarly (e.g. HD reconstruction with 2 iterations and 2 mm Gaussian smoothing).
机译:我们描述了对对比度的对比度与噪声,精度与精度的精度Vs噪声,精度与Siemens的沟通时间飞行时间(TOF)PET扫描仪的互相差异进行评估。 HD-PET重建是具有点扩散功能(PSF)建模的3D-OSEM,HD + TOF是具有时间扩展功能(TSF)建模的HD-PET,基于从中获取的TOF信息扫描器。扫描胸脊幻影,使用HD和HD + TOF方法重建数据。作为图像中的体素噪声的函数评估热对比度以及从具有不同密度的区域的冷对比。对比度的变异系数(COV)在低统计中评估了10多个实现,并绘制为相反偏差的函数。还比较HD和HD + TOF重建之间的标准偏差(STD)图像。观察到HD + TOF重建在热对比度VS噪声方面优于HD方法,以及低密度区域中的冷对比度。然而,对于高密度区域,与HD方法相比,HD + TOF显示出略差差。在精度VS精度方面也观察到一致的结果。 STD图像比较显示了HD和HD + TOF重建之间的类似结果。另外,幻像数据也被用于传记-MCT(MCT)和传记-40(B40)的各种方法和参数,并且对比度与噪声评估显示相同的重建方法在两个扫描仪上具有相同或相似的参数同样执行(例如,使用2次迭代和2 mm高斯平滑的HD重建)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号