We describe an evaluation of the HD and HD+TOF reconstructions in terms of contrast vs noise, precision vs accuracy, and inter-realization variation on Siemens' Biograph-mCT time-of-flight (TOF) PET scanner. The HD-PET reconstruction is the 3D-OSEM with point-spread-function (PSF) modeling, and the HD+TOF is the HD-PET with the time-spread-function (TSF) modeling based on the TOF information acquired from the scanner. A chest-spine phantom was scanned, and the data were reconstructed using the HD and HD+TOF methods. Hot contrast as well as cold contrast from regions with different densities was evaluated as a function of the voxel-noise in the images. The coefficient of variation (COV) in contrast was evaluated over 10 realizations at low statistics and plotted as a function of the bias in contrast. The standard deviation (STD) images were also compared between the HD and HD+TOF reconstructions. It was observed that the HD+TOF reconstruction outperforms the HD method in terms of hot contrast vs noise as well as cold contrast in the low density regions. However, for the high density regions the HD+TOF showed slightly worse results as compared to the HD method. Consistent results were also observed in terms of precision vs accuracy. The STD image comparison showed similar results between the HD and HD+TOF reconstructions. Additionally, the phantom data were also reconstructed with various methods and parameters for both Biograph-mCT (mCT) and Biograph-40 (B40), and the contrast vs noise evaluation showed that the same reconstruction method with the same or similar parameters on both scanners performs similarly (e.g. HD reconstruction with 2 iterations and 2 mm Gaussian smoothing).
机译:我们将在西门子的Biograph-mCT飞行时间(TOF)PET扫描仪上描述对比度和噪声,精度与准确性以及帧间实现差异方面对HD和HD + TOF重构的评估。 HD-PET重建是具有点扩展功能(PSF)建模的3D-OSEM,而HD + TOF是基于时间扩展功能(TSF)建模的HD-PET,其基于从TFP信息中获取的TOF信息。扫描器。扫描胸部-脊柱体模,并使用HD和HD + TOF方法重建数据。来自不同密度区域的热对比度和冷对比度被评估为图像中体素噪声的函数。对比的变异系数(COV)在低统计量下通过10个实现进行了评估,并绘制为对比偏差的函数。还比较了HD和HD + TOF重建之间的标准差(STD)图像。观察到,在低密度区域中,热对比度,噪声以及冷对比度方面,HD + TOF重建优于HD方法。但是,对于高密度区域,与HD方法相比,HD + TOF的结果稍差。在精度与准确性方面也观察到一致的结果。 STD图像比较显示了HD和HD + TOF重建之间的相似结果。此外,还使用各种方法和参数为Biograph-mCT(mCT)和Biograph-40(B40)重建了幻像数据,并且对比度与噪声评估表明,在两个扫描仪上具有相同或相似参数的相同重建方法表现类似(例如,具有2次迭代和2 mm高斯平滑的高清重建)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号