Resolution and noise properties of 123I MIBG SPECT with collimator-detector response modeling

机译:具有准直器响应响应建模 123 I MIBG SPECT的分辨率和噪声属性

获取原文

摘要

The effects of collimator-detector response (CDR) were studied for two clinical SPECT systems, the Siemens Symbia and GE Infinia, in terms of image resolution, noise properties, and activity estimation of 123I-MIBG in the heart. For each camera, Monte Carlo simulations were used to create CDR models, which were incorporated into a rotation-based projector for 3D-OS-EM SPECT reconstruction. The CDR models were used to generate simulated projections to study the degrading effects of the various CDR components, as well as the efficacy of compensating these components. In simulated images contaminated by 123I septal penetration and detector scatter, compensating the geometric response alone qualitatively restored image contrast, however, activity in the heart was significantly underestimated. The CDR models and the effect of CDR compensation were highly system-dependent. The contribution of detector scatter to activity underestimation was 8–10% for the Symbia and 18–20% for the Infinia, while the contribution of septal penetration was 30% for the Symbia and 5% for the Infinia. The simulation results were confirmed in SPECT measurements of a physical torso phantom on a Symbia SPECT/CT system. These results demonstrate that accurate activity estimation of cardiac 123IMIBG SPECT may require CDR compensation of septal penetration and detector scatter.
机译:在图像分辨率,噪声属性和 123 i-mibg的图像分辨率,噪声属性和活动估计方面,研究了两种临床Symbia和Ge Infinia的两种临床Symbia和Ge Infinia的疗效。心脏。对于每个摄像头,Monte Carlo模拟用于创建CDR模型,该模型被纳入基于旋转的投影仪,用于3D-OS-EM SPECT重建。 CDR模型用于生成模拟突起以研究各种CDR组分的降低效果,以及补偿这些组分的功效。在由 123 i隔间渗透和探测器散射污染的模拟图像中,单独补偿单独的几何响应恢复的图像对比度,然而,心脏中的活性显着低估了。 CDR模型和CDR补偿的效果高度系统依赖性。综合症状的探测器对活动的贡献为低估为8-10%,对于Infinia而言,18-20%,而Systal渗透的贡献为Symbia的30%,而Infinia为5%。在Symbia Spect / CT系统上的物理躯干幻像的SPECT测量中确认了模拟结果。这些结果表明,心脏 123 IMIBG SPECT的准确活性估计可能需要CDR补偿隔膜渗透和探测器散射。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号