首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >Spatial resolution uniformity, isotropy, and the effect of depth of interaction information in a 1mm3 resolution, limited-angle PET system
【24h】

Spatial resolution uniformity, isotropy, and the effect of depth of interaction information in a 1mm3 resolution, limited-angle PET system

机译:在1mm3分辨率的有限角度PET系统中,空间分辨率的均匀性,各向同性以及交互作用深度信息的影响

获取原文

摘要

Limited-angle tomography PET systems use configurations that result in incomplete and irregular sampling of projection data. In a two paneled, 1mm resolution PET system dedicated to breast imaging with 1mm depth-of-interaction (DOI) resolution, we measure both the uniformity and isotropy of the spatial resolution. Uniformity measures the variation in spatial resolution over the field of view (FOV), and isotropy measures the difference in resolution in the x, y, and z dimensions. Reconstructing data from point sources in air placed across the FOV with MLEM, we measure spatial resolution across the FOV in 3 dimensions. At the center of the FOV, the resolution was 0.775 mm, 1.296 mm, and 0.782mm in x, y, and z, respectively. We examine the impact of MLEM iterations on the measured uniformity and isotropy of spatial resolution. We compare MLEM resolutions to metrics in literature, see some correspondence, and discuss their limitations. We step a point source in 100 μm steps to show the ability of the system to resolve closely spaced points. The system separates point sources by more than one FWHM at 0.700mm spacing. DOI resolutions impact is studied by varying the binning resolution of interactions in crystals along the direction away from the region of interest. Resolution in all directions is substantially impacted by DOI rebinning. Resolution in the DOI direction (orthogonal to the panels) degrades from 1.296mm FWHM at 1mm DOI resolution to 5.908mm with 16mm DOI resolution showing the importance of DOI and sampling from oblique lines of response in limited-angle systems.
机译:有限角度断层扫描PET系统使用的配置会导致投影数据的采样不完整和不规则。在专门用于乳腺成像且具有1mm的相互作用深度(DOI)分辨率的两块面板,1mm分辨率的PET系统中,我们测量了空间分辨率的均匀性和各向同性。均匀性可测量视野(FOV)上空间分辨率的变化,而各向同性可测量x,y和z维度上分辨率的差异。使用MLEM重构从整个FOV放置的空气中的点源数据,我们在3个维度上测量了FOV的空间分辨率。在FOV的中心,x,y和z的分辨率分别为0.775 mm,1.296 mm和0.782mm。我们研究了MLEM迭代对所测得的空间分辨率的均匀性和各向同性的影响。我们将MLEM分辨率与文献中的指标进行比较,查看一些对应关系,并讨论它们的局限性。我们以100μm的步长步进一个点源,以显示系统解析间距较小的点的能力。该系统以0.700mm的间距通过多个FWHM分离点源。通过沿远离感兴趣区域的方向改变晶体中相互作用的装箱分辨率来研究DOI分辨率的影响。 DOI重组会严重影响所有方向的分辨率。 DOI方向(垂直于面板)的分辨率从1mm DOI分辨率的1.296mm FWHM降低到5.908mm(16mm DOI分辨率),这表明DOI的重要性以及在有限角度系统中从斜响应线采样。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号