首页> 外文会议>Conference on physics of medical imaging >Dependence of Radiation Dose on Area and Volumetric Mammographic Breast Density Estimation
【24h】

Dependence of Radiation Dose on Area and Volumetric Mammographic Breast Density Estimation

机译:辐射剂量对乳房面积和乳腺体积密度的估计

获取原文

摘要

Mammographic breast density is a strong risk factor for breast cancer. Studies on imaging dose in mammography have primarily focused on imaging quality and diagnostic accuracy, while little work has been done on understanding its effect on the estimation of breast density. Studies on the effect of dose on mammographic density estimation can be useful in dose reduction for the purpose of density estimation and monitoring. In this study, we investigate the dependence of percent area (PD%) and volumetric (VD%) breast density estimation on imaging dose using an anthropomorphic breast phantom (Rachel, Gammex). A set of digital mammograms were obtained with a GE Senographe 2000D FFDM system, using 220 unique combinations of different imaging physics, namely target/filter, kVp and mAs. Specifically, 8 different mAs settings were defined as corresponding to 10%, 20%, 40%, 70%, 100%, 150%, 200% and 300% of 1.8 mGy reference average glandular dose (AGD) for standard phototimed exposure. Breast density was estimated using fully-automated FDA-cleared software (Quantra v.2.0, Hologic Inc.). The obtained estimates were analyzed to study the effect of the imaging dose, using ANOVA and linear regression. Results show that there is a statistically significant dependence of density estimation on x-ray imaging dose (p-value=0.014 and <0.001 for PD% and VD%, respectively), while the actual variation of the estimation across the different levels of dose is relatively low (standard deviation of 2.87% and 0.66% for PD% and VD% respectively), the differences could be significant when breast density measures are used for risk estimation.
机译:乳腺钼靶X光检查是乳腺癌的重要危险因素。乳腺摄影中成像剂量的研究主要集中在成像质量和诊断准确性上,而在了解其对乳房密度估计的影响方面所做的工作很少。研究剂量对乳腺X线摄影密度估计的影响可用于减少剂量,以达到密度估计和监测的目的。在这项研究中,我们调查了拟人化的乳房幻影(Rachel,Gammex)在成像剂量上对百分比面积(PD%)和体积(VD%)乳房密度估计的依赖性。使用220种不同成像物理学的独特组合(即目标/滤光片,kVp和mAs),使用GE Senographe 2000D FFDM系统获得了一组数字化乳房X线照片。具体而言,定义了8种不同的mAs设置,分别对应于标准光定时曝光的1.8 mGy参考平均腺体剂量(AGD)的10%,20%,40%,70%,100%,150%,200%和300%。使用全自动的FDA认证软件(Quantra v.2.0,Hologic Inc.)估算乳房密度。使用ANOVA和线性回归分析获得的估计值,以研究成像剂量的影响。结果表明,密度估计值对X射线成像剂量有统计学意义的依赖性(PD%和VD%的p值分别为0.014和<0.001),而不同剂量水平的估计值的实际差异相对较低(PD%和VD%的标准偏差分别为2.87%和0.66%),当使用乳房密度测量方法进行风险估计时,差异可能会很大。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号