首页> 外文会议>SICE Annual Conference >Study on the reliability of comparison method for thermocouples in industry around 1324 #x00B0;C
【24h】

Study on the reliability of comparison method for thermocouples in industry around 1324 #x00B0;C

机译:1324°C左右产业热电偶比较方法可靠性研究

获取原文

摘要

To examine the reliability of thermocouple comparison in industry at high temperature, thermocouple comparison measurements were performed around 1324 °C in nine laboratories by using two Pt/Pd thermocouples and one type R thermocouple following calibration at Co-C eutectic point. It was found that the standard deviations of the emf values of the Pt/Pd-2 (one of the Pt/Pd thermocouples) and type R thermocouples among nine participating laboratories were 4.2 μV (0.2 °C) and 5.0 μV (0.4 °C), respectively.
机译:为了检查在工业中的热电偶比较的可靠性在高温下,通过在CO-C共晶点校准后使用两个Pt / Pd热电偶和一种型R热电偶在九个实验室中进行热电偶比较测量。发现Pt / PD-2(Pt / Pd热电偶之一)和九个参与实验室中的R热电偶的标准偏差为4.2μV(0.2°C)和5.0μV(0.4°C ), 分别。