首页> 外文会议>World Congress on Medical Physics and Biomedical Engineering >Detectability in SPECT Myocardial Perfusion Imaging: Comparison between a Conventional and a Semiconductor Detector System
【24h】

Detectability in SPECT Myocardial Perfusion Imaging: Comparison between a Conventional and a Semiconductor Detector System

机译:SPECT心肌灌注成像中的可检测性:传统和半导体探测器系统之间的比较

获取原文

摘要

The aim of this study is to compare cold lesions detectability in myocardial perfusion SPECT images acquired by two cameras: a conventional SPECT with scintillation detector (Philips Cardio MD) and a cadmium-zinc-telluride (CZT) semiconductor detector system(GE Discovery NM 530C). The test object was an anthropomorphic torso phantom with cardiac insert, filled with ~(99m)Tc, simulating a SPECT myocardial perfusion acquisition. Images were reconstructed using the standard clinical department protocol, and analyzed by axial and lateral count profiles, and polar distributions. The lateral profile showed that the contrast between the lesion region and background in the Cardio MD camera is higher (62.2%) than in the CZT camera (42.5%). The axial profile showed a higher contrast (37.0%) in the CZT camera compared to the Cardio MD (31.9%). CZT showed a higher contrast between background and the maximum activity point along the line (68.2%) than in the Cardio MD (57.5%). Performing the maximum relative count region and the lesion area analysis, it was verified that the contrast between these regions are 31.2% and 25.5% to CZT and Cardio MD, respectively. The polar map in the CZT camera displays a better uniformity between the segments' counting, which means the lesion identification is clearer than in the Cardio MD camera, where the distribution is diffuse and the lesion is unclear. The results showed that the cold myocardium lesion detectability is enhanced in the CZT camera. In conclusion, the physical performance of CZT camera is higher than conventional camera, but further studies are required to evaluate other parameters such as the influence of positioning and attenuation.
机译:该研究的目的是比较由两个相机获取的心肌灌注SPECT图像中的冷病变可检测性:与闪烁检测器(飞利浦Cardio MD)和镉 - 锌 - 碲化物(CZT)半导体探测器系统的传统SPECT(GE Discovery NM 530C )。测试对象是一种拟人躯干幻影,具有心脏插入物,填充〜(99m)TC,模拟SPECT心肌灌注采集。使用标准临床部门协议重建图像,并通过轴向和横向计数曲线和极性分布分析。横向轮廓显示,心脏MD相机中的病变区和背景之间的对比度高于CZT相机(42.5%)。与心脏MD(31.9%)相比,CZT相机中的轴向型材显示出更高的对比度(37.0%)。 CZT在沿线(68.2%)之间的背景和最大活动点之间的背景和最大活动点(57.5%)显示出更高的对比度。执行最大相对计数区域和病变区分析,验证了这些区域之间的对比度分别为CZT和Cardio MD的31.2%和25.5%。 CZT摄像机中的极性图在段计数之间显示更好的均匀性,这意味着病变识别比有氧MD摄像机更清晰,其中分布是漫射的,并且病变尚不清楚。结果表明,CZT相机中的耐寒心肌病变可检测性得到增强。总之,CZT相机的物理性能高于传统摄像机,但需要进一步的研究来评估其他参数,例如定位和衰减的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号