Assuring consistency of respiratory resistance measurements

机译:确保呼吸阻力测量的一致性

获取原文

摘要

We have developed a simple pulmonary diagnostic device that measures respiratory resistance of adults and children. The device is Airflow Perturbation Device (APD). There are many commercially available respiratory diagnostic devices in clinical use, but the major advantage of the APD is that it is an effortless device that evaluates the respiratory resistance effortlessly. Further, it can do this only by normal breathing through the device for less than one minute. To assure the consistency of the respiratory resistance values obtained using the APD, we designed a system that simulates the normal range of human breathing. We further compared the APD readings with a clinically accepted respiratory diagnostic system, the Impulse Oscillometry (IOS). IOS is an FDA approved system that also, similar to APD, measures respiratory resistance. The measurements using the APD and IOS were conducted at constant volumetric flow rate of 0.15 L·s−1 1. The results show that the resistance values obtained from the APD and IOS were consistent and linearly correlated (R2=0.999), we were then assured that the respiratory resistance values obtained using the APD are valid and consistent.
机译:我们已经开发出一种简单的肺部诊断设备,可以测量成人和儿童的呼吸阻力。该设备是气流扰动设备(APD)。临床上有许多可商购的呼吸诊断设备,但是APD的主要优点是它是一种毫不费力的设备,可以毫不费力地评估呼吸阻力。此外,它只能通过正常呼吸通过设备少于一分钟来执行此操作。为了确保使用APD获得的呼吸阻力值的一致性,我们设计了一个模拟人的正常呼吸范围的系统。我们进一步将APD读数与临床认可的呼吸诊断系统即脉冲示波法(IOS)进行了比较。 IOS是FDA批准的系统,它也类似于APD,可以测量呼吸阻力。使用APD和IOS进行的测量是在0.15 L·s -1 1的恒定体积流量下进行的。结果表明,从APD和IOS获得的电阻值是一致且线性相关的(R 2 = 0.999),然后我们确信使用APD获得的呼吸阻力值是有效且一致的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号