首页> 外文期刊>Journal of Medical Engineering >Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
【24h】

Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry

机译:气流微扰装置与脉冲示波法测量呼吸阻力的比较

获取原文
           

摘要

The airflow perturbation device (APD) has been developed as a portable, easy to use, and a rapid response instrument for measuring respiratory resistance in humans. However, the APD has limited data validating it against the established techniques. This study used a mechanical system to simulate the normal range of human breathing to validate the APD with the clinically accepted impulse oscillometry (IOS) technique. The validation system consisted of a sinusoidal flow generator with ten standardized resistance configurations that were shown to represent a total range of resistances from 0.12 to 0.95 kPa·L−1·s (1.2–9.7 cm H2O·L−1·s). Impulse oscillometry measurements and APD measurements of the mechanical system were recorded and compared at a constant airflow of 0.15 L·s−1. Both the IOS and APD measurments were accurate in assessing nominal resistance. In addition, a strong linear relationship was observed between APD measurements and IOS measurements (R2 = 0.999). A second series of measurements was made on ten human volunteers with external resistors added in their respiratory flow paths. Once calibrated with the mechanical system, the APD gave respiratory resistance measurements within 5% of IOS measurements. Because of their comparability to IOS measurements, APD measurements are shown to be valid representations of respiratory resistance.
机译:气流扰动设备(APD)已开发为一种便携式,易于使用的快速响应仪器,用于测量人体的呼吸阻力。但是,APD的数据有限,无法根据既定技术对其进行验证。这项研究使用机械系统模拟人的正常呼吸,以通过临床认可的脉冲示波法(IOS)技术验证APD。验证系统由具有十种标准化电阻配置的正弦流发生器组成,这些电阻配置显示出的总电阻范围为0.12至0.95 kPa·L-1·s(1.2–9.7 cm H2O·L-1·s)。记录并记录了机械系统的脉冲示波法测量值和APD测量值,并在0.15 L·s-1的恒定气流下进行了比较。在评估标称电阻时,IOS和APD的测量均准确。另外,在APD测量值和IOS测量值之间观察到了很强的线性关系(R2 = 0.999)。在十名人类志愿者的呼吸流动路径中添加了外部电阻器之后,进行了第二系列的测量。一旦通过机械系统进行了校准,APD的呼吸阻力测量值将在IOS测量值的5%之内。由于它们与IOS测量值具有可比性,因此显示APD测量值是呼吸阻力的有效代表。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号