Image quality fig. evaluator used in radiography based on contrast-detail phantom

机译:图像质量基于对比细节体模的射线照相评估仪

获取原文

摘要

In radiology, it is significantly important to produce adequate diagnostic information for affecting the patient with the lowest amount of dose. A contrast-detail phantom is generally used to study the quality of the image and the amount of radiation dose for a digital X-ray imaging system. For evaluating the quality of a phantom image, the radiologists are required to indicate the location of the holes in each square in the phantom image. Then, the image quality figure (IQF) of the image can be calculated. However, evaluation by the human eye is subjective and time-consuming. In this paper, an image processing based IQF evaluation method is proposed to automatically measure the quality of a phantom image. The experimental results tell that the proposed method is more sensitive in estimating the IQF of a phantom image than the observation of radiologists.
机译:在放射学中,产生足够的诊断信息以影响剂量最低的患者非常重要。通常使用对比度细节模型来研究数字X射线成像系统的图像质量和辐射剂量。为了评估体模图像的质量,要求放射线医师指出体模图像中每个正方形中的孔的位置。然后,可以计算图像的图像质量指数(IQF)。但是,人眼评估是主观且费时的。本文提出了一种基于图像处理的IQF评估方法来自动测量幻像图像的质量。实验结果表明,与放射科医生的观察相比,所提出的方法在估计幻像图像的IQF方面更为敏感。