The Radiometric Calibration Facility (RCF) at Northrop Grumman Aerospace Systems was used between 1995 and 2008 to establish the pre-launch calibration of the first six Clouds and the Earth's Radiant Energy System (CERES) instruments, with the seventh CERES instrument scheduled to be tested in the RCF in 2012. This paper summarizes the performance of the RCF radiometric standards NFBB (Narrow-field Black-body) and SWRS (Short-wave Reference Source), as well as the RCF TACR (Transfer Active Cavity Radiometer), in the context of the CERES pre-launch calibration process. A detailed investigation of the long-term stability of these standards using 1999–2008 CERES FM5 testing is summarized, showing the RCF sources in the long-wave and short-wave have remained stable relative to the NFBB absolute reference at levels of 0.06% and 0.5%, respectively, over the 8-year span.
机译:1995年至2008年间,诺斯罗普·格鲁曼航空系统公司的辐射校准设施(RCF)用于建立前六颗云和地球辐射能系统(CERES)仪器的发射前校准,并计划对第七台CERES仪器进行测试本文概述了RCF辐射标准NFBB(窄场黑体)和SWRS(短波参考源)以及RCF TACR(传输有源腔辐射计)在2012年的RCF中的性能。 CERES启动前校准过程的背景。总结了使用1999–2008 CERES FM5测试对这些标准的长期稳定性进行的详细调查,结果表明,长波和短波中的RCF源相对于NFBB绝对参考值都保持稳定,为0.06%,而在8年的时间里分别为0.5%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号