【24h】

Quality controls in scanning test rig for air filters

机译:空气滤清器扫描测试台中的质量控制

获取原文

摘要

This paper describes the result of studies on quality control in a scanning test rig for air filters. Measurements were carried out including the accuracy of flowrate sampling, the performance of dilutor, the stability of aerosol generator, the uniformity of aerosol in the cross sections, the background concentration at the downstream of filters, the characteristics of motion components, the response time of sampling, and the ability to detect leakage on filters. Results were compared with the requirement described by EN 1822. It was shown that the scanning test rig is suitable for pressure drop, efficiency and leakage test of HPEA filters.
机译:本文介绍了在空气过滤器扫描测试台上进行质量控制的研究结果。进行的测量包括流量采样的准确性,稀释器的性能,气溶胶发生器的稳定性,横截面中气溶胶的均匀性,过滤器下游的背景浓度,运动部件的特性,响应时间等。采样,以及检测过滤器泄漏的能力。将结果与EN 1822中描述的要求进行了比较。结果表明,扫描测试台适用于HPEA过滤器的压降,效率和泄漏测试。