Dynamic chest radiography with a flat-panel detector (FPD):Ventilation-perfusion study

机译:带有平板探测器(FPD)的动态胸部放射线照相术:通气灌注研究

获取原文
获取外文期刊封面目录资料

摘要

Pulmonary ventilation and blood flow are reflected in dynamic chest radiographs as changes in X-ray translucency, i.e., pixel values. This study was performed to investigate the feasibility of ventilation-perfusion (V/Q) study based on the changes in pixel value. Sequential chest radiographs of a patient with ventilation-perfusion mismatch were obtained during respiration using a dynamic flat-panel detector (FPD) system. The lung area was recognized and average pixel value was measured in each area, tracking and deforming the region of interest. Inter-frame differences were then calculated, and the absolute values were summed in each respiratory phase. The results were visualized as ventilation, blood flow, V/Q ratio distribution map and compared to distribution of radioactive counts on ventilation and perfusion scintigrams. In the results, abnormalities were appeared as a reduction of changes in pixel values, and a correlation was observed between the distribution of changes in pixel value and those of radioactivity counts (Ventilation; r=0.78, Perfusion; r=0.77). V/Q mismatch was also indicated as mismatch of changes in pixel value, and a correlation with V/Q calculated by radioactivity counts (r=0.78). These results indicated that the present method is potentially useful for V/Q study as an additional examination in conventional chest radiography.
机译:肺通气和血流量在动态胸部X线照片中反映为X射线半透明性(即像素值)的变化。进行这项研究是为了研究基于像素值变化的通气灌注(V / Q)研究的可行性。使用动态平板检测器(FPD)系统在呼吸过程中获得了通气-灌注不匹配的患者的连续胸部X线照片。识别出肺部区域,并在每个区域中测量平均像素值,从而跟踪并变形感兴趣区域。然后计算帧间差异,并对每个呼吸阶段的绝对值求和。将结果可视化为通气,血流量,V / Q比分布图,并与通气和灌注闪烁图上的放射性计数分布进行比较。结果,异常现象表现为像素值变化的减少,并且在像素值变化的分布与放射性计数的分布之间观察到相关性(通气; r = 0.78,灌注; r = 0.77)。 V / Q不匹配也表示为像素值变化的不匹配,并且与通过放射性计数计算的V / Q相关(r = 0.78)。这些结果表明,本方法对于V / Q研究潜在地有用,作为常规胸部X线照相术中的附加检查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号