首页> 外文会议>Conference on physics of medical imaging >K-edge subtraction imaging using a pixellated energy-resolving detector
【24h】

K-edge subtraction imaging using a pixellated energy-resolving detector

机译:使用像素化能量分辨探测器的K边缘减影成像

获取原文

摘要

This paper presents preliminary work aimed at assessing the feasibility of K-edge subtraction imaging using the spectroscopic information provided by a pixellated energy-resolving Cadmium Zinc Telluride detector, having an active area of 20x20 pixels 250 μm in size. Images of a test object containing different amounts of Iodine-based contrast agent were formed above and below the K-edge of Iodine (33.2 keV) by integrating, pixel by pixel, different windows of the spectrum. The results show that the optimum integration window for details 1-2 mm in diameter is between 2 keV and 5 keV. Concentrations of down to 50 μg Iodine/ml were detected in a 1-mm diameter tube with an entrance dose of 100 μGy.
机译:本文介绍了旨在使用像素化能量分辨碲化镉锌锌探测器提供的光谱信息评估K边缘相减成像的可行性的初步工作,该探测器的有效面积为20x20像素,尺寸为250μm。通过逐像素积分光谱的不同窗口,在碘的K边缘(33.2 keV)上方和下方形成了包含不同量的基于碘的造影剂的测试对象的图像。结果表明,直径为1-2 mm的细节的最佳积分窗口在2 keV和5 keV之间。在直径为1 mm的试管中检测到低至50μg碘/ ml的浓度,入口剂量为100μGy。