Verification of the performance accuracy of a real-time skin-dose tracking system for interventional fluoroscopic procedures

机译:验证用于介入荧光镜检查程序的实时皮肤剂量跟踪系统的性能准确性

获取原文

摘要

A tracking system has been developed to provide real-time feedback of skin dose and dose rate during interventional fluoroscopic procedures. The dose tracking system (DTS) calculates the radiation dose rate to the patient's skin using the exposure technique parameters and exposure geometry obtained from the x-ray imaging system digital network (Toshiba Infinix) and presents the cumulative results in a color mapping on a 3D graphic of the patient. We performed a number of tests to verify the accuracy of the dose representation of this system. These tests included comparison of system-calculated dose-rate values with ionization-chamber (6 cc PTW) measured values with change in kVp, beam filter, field size, source-to-skin distance and beam angulation. To simulate a cardiac catheterization procedure, the ionization chamber was also placed at various positions on an Alderson Rando torso phantom and the dose agreement compared for a range of projection angles with the heart at isocenter. To assess the accuracy of the dose distribution representation, Gafchromic film (XR-RV3, ISP) was exposed with the beam at different locations. The DTS and film distributions were compared and excellent visual agreement was obtained within the cm-sized surface elements used for the patient graphic. The dose (rate) values agreed within about 10% for the range of variables tested. Correction factors could be applied to obtain even closer agreement since the variable values are known in real-time. The DTS provides skin-dose values and dose mapping with sufficient accuracy for use in monitoring diagnostic and interventional x-ray procedures.
机译:已经开发出跟踪系统,以在介入荧光检查过程中提供皮肤剂量和剂量率的实时反馈。剂量跟踪系统(DTS)使用从X射线成像系统数字网络(Toshiba Infinix)获得的曝光技术参数和曝光几何来计算对患者皮肤的辐射剂量率,并在3D色彩映射中显示累积结果患者的图形。我们进行了许多测试,以验证该系统剂量表示的准确性。这些测试包括将系统计算的剂量率值与电离室(6 cc PTW)的测量值进行比较,并随kVp的变化,射束滤波器,场大小,源到皮肤距离和射束角度进行比较。为了模拟心脏导管插入术,将电离室也放置在Alderson Rando躯干体模上的各个位置,并比较剂量投影与心脏等角点范围内的投影角度范围。为了评估剂量分布表示的准确性,用光束在不同位置曝光了Gafchromic膜(XR-RV3,ISP)。比较了DTS和胶片的分布,并在用于患者图形的cm大小的表面元素内获得了出色的视觉一致性。对于测试的变量范围,剂量(速率)值在约10%的范围内一致。由于变量值是实时已知的,因此可以应用校正因子以获得更紧密的一致性。 DTS提供足够的准确皮肤剂量值和剂量图,用于监测诊断和介入性X射线过程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号