摘要

Most phantom models do not include distinct definitions of tumor and normal tissue. This article proposes a novel solid breast - tumor phantom for microwave hyperthermia in breast cancer treatment. According to tissue dielectric properties reported, we performed phantom. A Network Analyzer and Dielectric Probe Kit measured dielectric constant and electrical conductivity for both breast normal tissue and tumor phantom. Measurements were through at frequency range from 2 to 3 GHz due microwave operation frequency is usually 2.45 GHz. At 2.45 GHz breast phantom had an electrical conductivity of 0.1304±0.0680 S/m and a dielectric constant of 4.4401 ± 1.2514. At the same frequency, tumor phantom had an electrical conductivity of 2.7015 ± 0.0783 S/m and a dielectric constant of 55.2566 ± 1.1899. The presented results highly correlate with the expected values. Therefore this phantom can be used for microwave hyperthermia in breast cancer treatment and antenna performance analysis and optimization.
机译:大多数体模模型不包括肿瘤和正常组织的明确定义。本文提出了一种新型的实体乳腺肿瘤模型,用于在乳腺癌治疗中进行微波热疗。根据报道的组织介电特性,我们进行了幻像。网络分析仪和介电探针套件可测量乳腺正常组织和肿瘤体模的介电常数和电导率。由于微波工作频率通常为2.45 GHz,因此在2至3 GHz的频率范围内进行了测量。在2.45 GHz时,胸模的电导率为0.1304±0.0680 S / m,介电常数为4.4401±1.2514。在相同的频率下,肿瘤体模的电导率为2.7015±0.0783 S / m,介电常数为55.2566±1.1899。给出的结果与预期值高度相关。因此,该模型可用于乳腺癌治疗中的微波热疗以及天线性能分析和优化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号