Asymmetrical oscillometric pulse waveform envelopes in normotensive and hypertensive subjects

机译:血压正常和高血压受试者的非对称示波脉冲波形包络

获取原文

摘要

This study investigated the symmetry of the oscillometric waveform shape obtained during noninvasive auscultatory blood pressure measurement. The oscillometric cuff pressures were recorded from ten normotensive and ten hypertensive subjects. The cuff pressure corresponding to the maximum oscillometric pulse was taken as the automated mean arterial pressure (MAP). The cuff pressures corresponding to 30%, 50% and 70% of the maximum oscillometric pulse peak amplitude were measured for both the high and low pressure regions. At the three different thresholds, the cuff pressure widths (absolute difference to the MAP) in the high and low pressure regions were compared. For both the normotensive and hypertensive groups, the cuff pressure widths in the high pressure region were significantly larger than in the low pressure region at the normalised amplitude levels of 50% and 70% (all P<;0.01), which confirmed an asymmetrical feature of the oscillometric waveform envelope.
机译:这项研究调查了在无创听诊血压测量过程中获得的示波波形形状的对称性。记录了十名血压正常和十名高血压受试者的示波法袖带压。将对应于最大示波脉搏的袖带压作为自动平均动脉压(MAP)。在高压区和低压区均测量了对应于最大示波脉冲峰值幅度的30%,50%和70%的袖带压力。在三个不同的阈值处,比较了高压区域和低压区域的袖带压力宽度(与MAP的绝对差)。对于正常血压组和高血压组,在标准幅度水平分别为50%和70%时,高压区域的袖带压力宽度明显大于低压区域的袖带压力宽度(所有P <; 0.01),这证实了不对称特征示波波形包络线。