Efficient arbitrary volume reslicing for Pre-scan-converted volume in an ultrasound backend

机译:超声后端中用于预扫描转换后体积的高效任意体积切片

获取原文

摘要

Arbitrary volume reslicing (AVR) is used to generate 2D planes of any orientation from 3D data. Pre-scan-converted (pre-SC) AVR avoids the 3D scan conversion (SC) step and directly uses a pre-SC volume to generate 2D AVR planes. AVR is typically performed on a PC system external to the ultrasound frontend and backend. In this paper, we present an efficient pre-SC AVR algorithm in an ultrasound backend system. The backend supports AVR natively with other backend functions, so the PC system does not need to compute 2D AVR planes from 3D volumes. We used phantom and in vivo datasets to evaluate the pre-SC and post-SC AVR. The normalized root mean squared difference (NRMSD) was 1.23% for phantom data, and 1.21% for in vivo data. Using one of 8 cores in a multi-core processor, Cell, our algorithm can generate a 1024×768 16-bit AVR image in 30.2 ms from a 512×136×64 16-bit input volume.
机译:任意体积重新弹性(AVR)用于生成3D数据的任何方向的2D平面。预先扫描转换(Pre-SC)AVR避免了3D扫描转换(SC)步骤,并直接使用预先使用预扫描卷来生成2D AVR平面。 AVR通常在超声前端和后端外部的PC系统上执行。在本文中,我们在超声后端系统中提出了一种高效的Pre-SC AVR算法。后端支持AVR与其他后端功能,因此PC系统不需要从3D卷中计算2D AVR平面。我们使用幻像和体内数据集来评估Pre-SC和后SC AVR。常规的根均方平方差(NRMSD)对于幻象数据为1.23%,体内数据为1.21%。在多核处理器中使用8个核心中的一个,我们的算法可以在30.2毫秒中生成1024℉的16位AVR图像,从512毫秒到512毫秒。 Ã,64 16位输入音量。