首页> 外文会议>International Workshop on Breast Imaging >Experimental dose estimation in anthropomorphic breast phantom in clinical mammography systems
【24h】

Experimental dose estimation in anthropomorphic breast phantom in clinical mammography systems

机译:临床乳腺X线摄影系统中拟人化乳房幻像的实验剂量估算

获取原文
获取外文期刊封面目录资料

摘要

The aim of this work is to estimate the percentage depth dose (PDD) and the mean glandular dose (MGD) at anthropomorphic breast phantoms using calibrated TLDs. For this task the TLD-100H was initially selected and calibrated in terms of air kerma, using a Radcal ionization chamber (IC). The experimental procedure was performed at a mammograph Mammomat 3000 Nova, located at the CDTN/CNEN facilities and all the detectors were exposed with 28 kV with distinct anode/filter combination (Mo/Mo, Mo/Rh and W/Rh). Furthermore, the TLDs were placed at the surface of anthropomorphic compressed breast phantoms (36-85 mm) and at 1,2 cm for depth doses measures, for PDD analyses. The MGD were estimated from entrance surface doses and using Dance's method. The dose-response curves for the TLDs indicated a good correlation coefficient (R~2 = 0,99) for all anode-filter combination with an uncertainty lower than 17%. The uncertainties of the measurements increased to maximum 35 % when the TLDs are placed the anthropomorphic breast phantom. The PDD were maximum 70%, at 1,2 cm, for the Mo/Mo target-filter combination. Regardless the target/filter (T/F) combination, the TL responses and consequently the MGDs substantially increased with the breast thickness. A maximum MGD of approximately 4.5 mGy was estimated for the 85 mm thickness breast, exposed to the Mo/Mo combination.
机译:这项工作的目的是使用校准的TLD估算拟人化的人体模型上的深度剂量百分比(PDD)和平均腺体剂量(MGD)。为此,最初选择了TLD-100H,并使用Radcal电离室(IC)根据空气比释动能进行了校准。实验程序在位于CDTN / CNEN设施的乳房X线摄影机Mammomat 3000 Nova上进行,所有检测器均暴露于28 kV的阳极/过滤器组合(Mo / Mo,Mo / Rh和W / Rh)。此外,将TLD放置在拟人化的压缩乳房模型(36-85毫米)的表面上,并放置在1,2厘米处,以进行深度剂量测量,以进行PDD分析。从入口表面剂量和使用Dance's方法估算MGD。 TLD的剂量反应曲线表明,所有阳极过滤器组合均具有良好的相关系数(R〜2 = 0.99),不确定度低于17%。当将TLD放置在拟人化的乳房体模上时,测量的不确定度增加到最大35%。对于Mo / Mo目标过滤器组合,PDD在1,2 cm处最大为70%。无论目标/过滤器(T / F)的组合如何,TL反应以及随之而来的MGD都会随着乳房厚度的增加而显着增加。对于暴露于Mo / Mo组合的85毫米厚的乳房,估计最大MGD约为4.5 mG​​y。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号