Purpose: To investigate the effect of age of mammography systems on the number of reported faults and equipment downtime. Method: Each screening unit in the NHS breast screening programmes logged any equipment faults in a centralised online fault reporting database. Data on faults occurring in 2018 for digital mammography systems were analysed. The relationship between the age of mammography systems and the number and consequences of equipment faults was examined. Severity of consequences was reported as number of days downtime, and number of cancelled appointments. Results: 2141 reported faults on 432 mammography systems were included in this analysis. The average age of an x-ray set at the time when a fault occurred was 5.76 years. 72% of mammography systems experienced five or fewer faults in 2018. A significant increase was observed in the number of faults and days of downtime reported on mammography systems six years old or greater, compared to those which were five years old or younger at the time the fault occurred (p<0.05 one-tailed, two-sample t-Test assuming unequal variances). A few mammography systems experienced a high number of faults and days downtime within their first year. The highest average number of reported faults and the highest severity of the consequences were found for mammography systems of around nine years old. Conclusions: The data presented could be used as evidence to support guidance on the age at which mammography systems used for breast screening should be replaced.
机译:目的:研究乳腺X线摄影系统的老化对报告的故障数量和设备停机时间的影响。方法:NHS乳房筛查程序中的每个筛查单元都会在集中的在线故障报告数据库中记录任何设备故障。分析了2018年数字化乳腺摄影系统发生的故障数据。检查了乳腺X线摄影系统的年龄与设备故障的数量和后果之间的关系。结果的严重程度报告为停机天数和已取消约会的数量。结果:这项分析包括432例乳腺X线摄影系统中2141例报告的故障。发生故障时设置的X射线平均年龄为5.76岁。在2018年,有72%的乳腺X线摄影系统发生了5个或更少的故障。与当时的5岁或以下的X射线摄影系统相比,发现6岁或以上的X射线摄影系统的故障数量和停机天数显着增加发生故障(假设方差不相等,p <0.05一尾两样本t检验)。少数X线乳腺摄影系统在使用的第一年就经历了大量的故障和几天的停机时间。对于九岁左右的乳腺X线摄影系统,发现的平均故障数最高,后果严重性最高。结论:所提供的数据可作为证据,为应该用于乳腺筛查的乳房X线照相系统应被更换的年龄提供指导。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号