首页> 外文会议>IEEE International Conference on Electronic Measurement and Instruments >Reasearch on performance of hard X-ray calibration device in low energy section
【24h】

Reasearch on performance of hard X-ray calibration device in low energy section

机译:硬X射线标定仪在低能段的性能研究

获取原文

摘要

The performance of hard X-ray test equipment is an important factor for ground calibration of satellite load detectors. In this paper, the low energy section $(8.3 sim 38.0$ keV) of hard X-ray single crystal test equipment using standard detector (high purity helium detector) Preliminary performance measurements were taken. It is measured that the energy linearity is 1 at $8.3 sim 38.0$ keV, the monochromatic light ratio is basically above 90%, the monochromatic luminous flux is substantially greater than 1000 cts/s, and the overall stability is about 0.6%.
机译:硬X射线测试设备的性能是卫星负载检测器地面校准的重要因素。在本文中,使用标准检测器(高纯度氦检测器)的硬X射线单晶测试设备的低能部分$(8.3 \ sim 38.0 $ keV)进行了初步性能测量。经测量,在$ 8.3 \ sim 38.0 $ keV处,能量线性为1,单色光比基本上高于90%,单色光通量实质上大于1000 cts / s,总体稳定性约为0.6%。