Research on Calibration Transfer Method of Portable Near Infrared Instrument

机译:便携式近红外仪器校准传递方法研究

获取原文

摘要

The establishment of the near-infrared spectrum prediction model is usually affected by environmental factors such as humidity and temperature, and other uncontrollable factors, which makes the original model unable to predict new sample components, resulting in a decline in model prediction accuracy. This paper discusses the calibration transfer method of TI-NANO portable near infrared instruments. The slope/intercept correction algorithm and the direct standardization correction algorithm are used for research. The results show that for the slope/intercept algorithm, the RMSEP of the model is reduced by 0.825, 0.509, 0.706, and 1.378 respectively, and R2 is increased by 0.133, 0.326, 0.116, and 0.288, respectively; for the direct standardization algorithm, the RMSEP of the model decreased by 0.547, 0.303, 0.897 and 1.633 respectively, and R2 increased by 0.195, 0.322, 0.135 and 0.204 respectively. After being processed by the calibration transfer method, the accuracy of the prediction results has been improved, especially the direct standardization method is preferable.
机译:近红外频谱预测模型的建立通常受环境因素的影响,例如湿度和温度,以及其他无法控制的因素,这使得原始模型无法预测新的样本组件,导致模型预测精度下降。本文讨论了TI-Nano便携式近红外仪器的校准传递方法。斜率/截距校正算法和直接标准化校正算法用于研究。结果表明,对于斜率/截距算法,模型的RMSEP分别减少0.825,0.509,0.706和1.378,R2分别增加0.133,0.326,0116和0.288;对于直接标准化算法,模型的RMSEP分别降低0.547,0.303,0.897和1.633,R2分别增加0.195,0.322,0.135和0.204。在通过校准传递方法处理之后,已经提高了预测结果的准确性,特别是直接标准化方法是优选的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号