Data-Based Parameter Calibration Method of the Computed Tomography System

机译:层析成像系统基于数据的参数标定方法

获取原文

摘要

This paper mainly studies data-based parameter calibration of the computed tomography (CT) system. By establishing a least squares algorithm model, CT system parameters were calibrated. First, the attenuation coefficient is calculated from the known data, and the relationship between the received information and the pitch of the X-ray transmission template is obtained by using the Lambert-beers principle. Then, by using the geometric relationship between the ray and the two-dimensional graph, the relationship model between the pitch and the variable to be sought is established. On this basis, we can get the multivariate linear equations which present the received information and the variables to be solved. Finally, the least square method is used to establish the target programming model which is based on the minimum square sum of the fitting value and the actual value of the received information. By using the genetic algorithm, the multiple linear equations are solved. Meanwhile, the deflection center and receiver unit spacing can be obtained.
机译:本文主要研究计算机断层扫描(CT)系统的基于数据的参数校准。通过建立最小二乘算法模型,对CT系统参数进行了校准。首先,从已知数据计算衰减系数,并且使用兰伯特-比尔斯原理获得接收信息与X射线透射模板的间距之间的关系。然后,通过使用射线和二维图之间的几何关系,建立音高和要寻找的变量之间的关系模型。在此基础上,我们可以得到表示所接收信息和待求解变量的多元线性方程。最后,使用最小二乘法建立目标编程模型,该模型基于拟合值和接收信息的实际值的最小平方和。通过使用遗传算法,求解了多个线性方程。同时,可以获得偏转中心和接收器单元的间隔。