首页> 外文会议>IEEE International Conference on Micro Electro Mechanical Systems >Ultra high-aspect-ratio neuroprobe: 5-μm-diameter and 400-μm-length needle detects action potentials in VIVO
【24h】

Ultra high-aspect-ratio neuroprobe: 5-μm-diameter and 400-μm-length needle detects action potentials in VIVO

机译:超高纵横比神经探针:直径5μm和长度400μm的针可检测VIVO中的动作电位

获取原文

摘要

This paper reports 5-μm-diameter and 400-μm-length ultra high-aspect-ratio needle electrodes for neuronal action potential recordings in vivo. The needle-electrode was vertically assembled on a 1 × 1 mm2 block module of silicon by vapor-liquid-solid (VLS) silicon growth technology for ~ 7 hours. After metallization and encapsulation of the needle, the module device was packaged for in vivo animal experiments. The fabricated needle-electrode, which has an electrical impedance of 328 kΩ at 1 kHz, shows the smooth tissue penetration and neural recording capabilities, as demonstrated by the needle penetration and recording action potentials of a mouse's cortex in vivo. To the author's knowledge, this is world's smallest extracellular recording needle, which will offer numerous electrophysiological recording applications while the brain tissue damage can be reduced. As a future work, the fabricated needle module device will further be minimized and arranged in an array of high dense needles with 100% device yield, while other functionalities (drug delivery and optogenetic applications) will be integrated with the same needle array.
机译:本文报道了直径为5μm和长度为400μm的超高长宽比针状电极,用于体内神经元动作电位的记录。通过汽-液-固(VLS)硅生长技术将针状电极垂直组装在1×1 mm2的硅块模块上约7小时。在对针进行金属化和封装之后,将模块设备包装起来用于体内动物实验。所制造的针电极在1 kHz时的电阻抗为328kΩ,显示出平滑的组织穿透和神经记录能力,如体内小鼠皮质的针穿透和记录动作电位所证明的那样。据作者所知,这是世界上最小的细胞外记录针,在可以减少脑组织损伤的同时,将提供多种电生理记录应用。作为未来的工作,将进一步最小化所制造的针头模块设备,并将其排列成高密度的针头阵列,从而实现100%的设备良率,而其他功能(药物输送和光遗传学应用)将与同一针头阵列集成在一起。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号