In Indonesia, the need for prosthetic knees is increasing due to the large number of amputees. However, the price of the prosthetic knee is rather expensive. Therefore, Biomechanics Research Team of ITB has been developing affordable prosthetic knee. This work presents the improvement of the previous design of the prosthetic knee which include the Design for Manufacture and Assembly (DFMA) analysis and static analysis. The calculated DFA index is 16.48% which means the assembly time is efficient. DFM analysis results indicate that cost to manufacture a prosthetic knee is around US$39. However, the design fails due to transfer of force from the lower body of the prosthetic knee to the linkages. Therefore, there is a need to modify the design and modification of the male pyramid adapter and the location of the adapter relative to the pylon are proposed. Static analysis shows that the maximum stress occurs on the snap ring of the prosthetic knee is 141.04 MPa, which is lower than its yield strength. The latest design has less number of components, obtained via DFMA analysis and also safe stress level.
机译:在印度尼西亚,由于截肢者数量众多,对假肢膝盖的需求正在增加。然而,假膝的价格相当昂贵。因此,ITB的生物力学研究团队一直在开发负担得起的假膝。这项工作提出了假肢膝关节先前设计的改进,其中包括制造和组装设计(DFMA)分析和静态分析。计算得出的DFA指数为16.48%,这意味着组装时间高效。 DFM分析结果表明,制造假肢的成本约为39美元。但是,由于将力从假肢的下半身传递到连杆,设计失败了。因此,需要修改凸棱锥适配器的设计和修改,并且提出适配器相对于塔架的位置。静态分析表明,假肢膝盖卡环上产生的最大应力为141.04 MPa,低于其屈服强度。最新设计的零件数量更少,可通过DFMA分析获得,并且应力水平也很安全。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号