首页> 外文会议>IET International Conference on Biomedical Image and Signal Processing >An optical fibre based ex-vivo device for detection of cytokines
【24h】

An optical fibre based ex-vivo device for detection of cytokines

机译:一种基于光纤的细胞因子的ex-体内装置

获取原文

摘要

Cytokine detection is an essential topic for biomedical research. Herein we demonstrated an ex-vivo device based on a silica optical fibre for monitoring of locally variable cytokine IL-6 concentrations using a sandwich immunoassay scheme. The fibre was designed to be introduced into the perforated catheter with micrometre size holes drilled along its length to enable fluid exchange between the outside and inside of the catheter. An exposed optical fibre (diameter 125 μm) modified with a layer of gold nanoparticles was functionalized with the IL-6 capture antibody to form the sensing interface. The IL-6 detection antibody which was loaded on the fluorescently labelled magnetic nanoparticles, was used for reporting the analyte signal. This device was successfully applied to detect cytokine IL-6 with the detection limit of 1 pg mL-1 in the sample volume of 1 μL. It has the linear detection range of 1-400 pg mL-1 and is capable of detecting localised IL-6 secreted by BV2 cells following their liposaccharide stimulation. This universal biological detection system is suitable as a platform for the in-vivo devices able to monitor multiple health conditions.
机译:细胞因子检测是生物医学研究的重要课题。在此,我们证明了基于二氧化硅光纤的前体内装置,用于使用夹层免疫测定方案监测局部可变细胞因子IL-6浓度。设计用于沿其长度钻的微型尺寸孔引入穿孔导管中以使导管的外部和内部之间的流体交换。用IL-6捕获抗体官能化用一层金纳米颗粒改性的暴露的光纤(直径125μm)以形成传感界面。将装载在荧光标记的磁性纳米颗粒上的IL-6检测抗体报告分析物信号。该装置成功地应用于检测细胞因子IL-6,检测极限为1μl样品体积的1pg ml-1。它具有1-400pg ML-1的线性检测范围,并且能够在脂质糖刺激后检测由BV2细胞分泌的局部IL-6。该通用生物检测系统适合作为能够监测多个健康状况的体内设备的平台。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号