首页> 外文会议>World biomaterials congress >Wear simulator testing of reverse total shoulder arthroplasty implants
【24h】

Wear simulator testing of reverse total shoulder arthroplasty implants

机译:反向全肩关节置换植入物的磨损模拟器测试

获取原文

摘要

Introduction: Reverse total shoulder arthroplasty (RTSA) retrieval humeral cups commonly exhibit damage in the inferior quadrant. Reported simulator wear rates range from 4-125 mm3/Mc using various motion and loading conditions. The present study assesses the wear of non-crosslinked RTSA implants using a custom wear simulation strategy, based in part on the load magnitudes and directions found in cadaveric RTSA reconstructed shoulder testing. This was expected to produce wear in the same locations as occurs in vivo and to help assess whether simulator wear rates could be high enough to cause clinical concerns regarding osteolysis. Methods: A modified orbital bearing hip wear simulator gave an adduction-abduction range of 30-97.5° and flexion-extension range of ±22.5°. A transient load profile with a 914 N peak was applied (Figure 1) similar to that observed in RTSA reconstructed cadaveric shoulders. Wear testing was performed at 1.13 Hz for 1.44 Mc on eight RTSA implants (DePuy Delta XTEND, size: 38 mm). Following previous knee simulator wear testing, the lubricant was bovine calf serum diluted with PBS to a 30 g/L protein concentration, with sodium hyaluronate (1.5 g/L) and an antibiotic-anrjmycotic (1%). Every 0.25 Mc, the specimens were cleaned and weighed. Results: All cups exhibited a large wear scar with a dull (non-reflective) scuffed appearance covering the entire intended articular surface, with only the superior-most surface remaining pristine (Figure 2). Four specimens exhibited a distinct thinner polished region near the superior wear scar margin. Inferior edge wear was present on all cups (red lines). Within the contact zone, all glenospheres exhibited light surface scratching, and after every 0.25 Mc there was some surface staining present, which was removed during ultrasonic cleaning. The wear rates of the humeral cups had considerable variation, with a range of 115 to 345 mm3/Mc (Figure 2), and a mean wear rate of 201 ± 87 mm3/Mc. Discussion: The present study had by far the highest rate compared to other wear studies, although there was little consensus regarding simulator conditions. No other studies incorporated HA in their lubricant, which has been shown to increase wear rate. An earlier pilot study used alpha calf serum and had a much lower wear rate, suggesting that the higher wear may be a consequence of the lubricant. The inferiomedial cup edge had considerable wear in the present simulator testing, similar to that found in clinical retrievals, damage which would have a detrimental effect on both contact mechanics and potentially the stability of the reconstructed joint in vivo. Conclusions: The high wear rate in the present study suggests that RTSA wear could become a long term clinical problem for the more active patients. Highly crosslinked polyethylene humeral cups may reduce wear but this substitution may exacerbate problems associated with scapular notching and thus new RTSA designs may also be needed.
机译:简介:反向全肩关节置换术(RTSA)取回肱骨杯通常在下象限表现出损伤。在各种运动和负载条件下,报告的模拟器磨损率范围为4-125 mm3 / Mc。本研究使用定制的磨损模拟策略评估非交联的RTSA植入物的磨损,部分基于尸体RTSA重建肩部测试中发现的载荷大小和方向。预期这会在体内发生的相同位置产生磨损,并有助于评估模拟器的磨损率是否可能高到足以引起有关溶骨的临床问题。方法:改进的轨道轴承髋关节磨损模拟器的内收外展范围为30-97.5°,屈伸范围为±22.5°。施加了一个具有914 N峰值的瞬态载荷曲线(图1),类似于在RTSA重建的尸体肩部观察到的。在八个RTSA植入物(DePuy Delta XTEND,尺寸:38 mm)上以1.13 Hz对1.44 Mc进行了磨损测试。在先前的膝盖模拟器磨损测试之后,润滑剂是用PBS将牛小牛血清稀释至30 g / L的蛋白质浓度,其中含有透明质酸钠(1.5 g / L)和抗生素-抗真菌药(1%)。每0.25 Mc,将样品清洗并称重。结果:所有的杯子都表现出较大的磨损痕迹,并在整个预期的关节表面上形成了钝化(非反射性)的磨损外观,只有最上面的表面保持原始状态(图2)。四个标本在卓越的磨损痕迹边缘附近显示出明显的较薄抛光区域。所有杯子(红线)的边缘磨损均较差。在接触区内,所有的小球体都表现出轻微的表面刮擦,每0.25 Mc后表面存在一些污点,在超声清洗过程中将其清除。肱骨杯的磨损率变化很大,范围为115至345 mm3 / Mc(图2),平均磨损率为201±87 mm3 / Mc。讨论:尽管在模拟器条件方面尚无共识,但本研究相比其他磨损研究具有最高的研究率。没有其他研究将HA掺入润滑剂中,已证明会增加磨损率。较早的一项先期研究使用了小牛血清,磨损率低得多,这表明较高的磨损可能是润滑剂造成的。在当前的模拟器测试中,下内侧杯缘具有相当大的磨损,类似于在临床取回中发现的那样,损坏将对接触力学和体内重建关节的稳定性均产生不利影响。结论:本研究中较高的磨损率表明RTSA的磨损可能成为较活跃的患者的长期临床问题。高度交联的聚乙烯肱骨杯可以减少磨损,但是这种替换可能会加剧与肩骨切口相关的问题,因此也可能需要新的RTSA设计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号