首页> 外文会议>International Conference on Biomedical Engineering >Design of Mass-Sensitive Sensor Array for Biomedical Application: Sensing Lipoproteins
【24h】

Design of Mass-Sensitive Sensor Array for Biomedical Application: Sensing Lipoproteins

机译:生物医学应用质量敏感传感器阵列的设计:感应脂蛋白

获取原文

摘要

Herein we report the design of a novel sensor array based on molecularly imprinted polymer (MIP) for low-density lipoprotein (LDL) and high-density lipoprotein (HDL) determination. MIP of each lipoprotein class comprising comonomer of methacrylic acid (MAA) and vinyl pyrrolidone (NVP) in the ratio of 3:2 and 2:3, were screened onto threeelectrode quartz crystal microbalance (QCM) as selective elements toward LDL and HDL, respectively. Sensor array device reveals concentration-dependent sensor responses toward varying amount of each standard LDL and HDL solution with a linear correlation coefficient of R2 = 0.89 and 0.99, respectively. HDL-MIP presents cross-reactivity to LDL at 9-19% compared to the LDL-MIP signal. LDL-MIP also has cross-reactivity to HDL at 0-3% of LDL-MIP responses. All sensor signals are fully reversible and the sensor array can be reused.
机译:在此,我们报告了基于用于低密度脂蛋白(LDL)和高密度脂蛋白(HDL)测定的分子印迹聚合物(MIP)的新型传感器阵列的设计。包括甲基丙烯酸(MAA)和乙烯基吡咯烷酮(NVP)的共聚单体的每个脂蛋白类的MIP,以3:2和2:3的比例筛选到三电极石英晶体微稳定(QCM)上,分别为LDL和HDL的选择性元素。传感器阵列装置揭示浓度依赖性传感器响应各标准LDL和HDL溶液的变化量,具有R的线性相关系数 2 分别= 0.89和0.99。与LDL-MIP信号相比,HDL-MIP在9-19%的情况下向LDL提供交叉反应性。 LDL-MIP还具有0-3%LDL-MIP响应的HDL的交叉反应性。所有传感器信号都是完全可逆的,传感器阵列可以重复使用。