首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Stability of Colorimetric Results in the Detection of Urine Biomarkers using a Paper-based Analytical Device
【24h】

Stability of Colorimetric Results in the Detection of Urine Biomarkers using a Paper-based Analytical Device

机译:比色度的稳定性导致使用纸的分析装置检测尿生物标志物

获取原文

摘要

Urinalysis allows obtaining information about many of the body's functions. Normally, urine is a very easily collected specimen, but in some cases, it can be an uncomfortable process. Moreover, standard methods such as microscopic examination or urine culturing represent time-consuming procedures. In this regard, this work explores the design of a device that combines disposable diapers, as a platform for sample collection, with part of the dipsticks method. Even though this is a desirable combination for a faster and comfortable approach, it is also convenient to extend the prescribed time, provided by the dipstick manufacturer, to read the results. This allows the stability and validity of the results for longer, preferably until the moment of the diaper changing, and it was also considered in the design. The desired conditions were reproduced for the reagent pads of glucose and nitrite. The results show that the RGB profile, obtained in glucose detection, varied after approximately 60 min. However, when the developed colors are visually compared with the reference colors, the results suggest that there were not significant variations. In the nitrite assay, by analyzing the RGB profile or by visual comparison with the reference colors, the color does not appear to change over time, but the results were not the expected. It seems that the proposed design is promising regarding results stability, but more experiments need to be done to check if the variations are constant for more biomarkers in the conditions provided by the device.
机译:尿液分析允许获得有关身体功能中的许多功能的信息。通常,尿液是一个非常容易收集的标本,但在某些情况下,它可能是一个不舒服的过程。此外,如微观检查或尿液培养的标准方法代表耗时的程序。在这方面,这项工作探讨了将一次性尿布组合的设备的设计,作为样品收集的平台,其中具有一部分的Dipsticks方法。即使这是一种更令人满意的组合,即较快且舒适的方法,它也方便地扩展了由Dipstick制造商提供的规定时间来阅读结果。这允许结果的稳定性和有效性更长,优选地直到尿布变化的那一刻,并且还在设计中考虑。为葡萄糖和亚硝酸盐的试剂垫再现所需的条件。结果表明,在葡萄糖检测中获得的RGB型材在大约60分钟后变化。然而,当显影色视到视觉上与参考颜色进行比较时,结果表明存在没有显着的变化。在亚硝酸盐测定中,通过分析RGB型材或通过与参考颜色进行视觉比较,颜色随着时间的推移似乎没有变化,但结果不是预期的。似乎所提出的设计是有关结果稳定性的承诺,但需要进行更多的实验以检查变型是否在设备提供的条件中的更多生物标志物恒定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号