首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Photoplethysmographic and SpO_2 Readings from the Neonatal Anterior Fontanelle: a case Study
【24h】

Photoplethysmographic and SpO_2 Readings from the Neonatal Anterior Fontanelle: a case Study

机译:来自新生儿前Fontanelle的PhotopleThysmography和Spo_2读数:一个案例研究

获取原文
获取外文期刊封面目录资料

摘要

There is a need for a more reliable, noninvasive and alternative measurement site for the monitoring of arterial blood oxygen saturation in critically ill neonates when peripheral perfusion is poor. The anterior fontanelle, a unique anatomical feature of the neonate, has been presented as an alternative site for the estimation of arterial blood oxygen saturation (SpO_2). A new fontanelle photoplethysmographic sensor and processing system has been developed to investigate fontanelle photoplethysmographic (PPG) signals and estimate SpO_2 values at this anatomical location. Preliminary clinical trials have shown that good quality PPG signals with large amplitudes and high signal to noise ratio can be obtained from the neonatal fontanelle. The estimation of SpO_2 values from the fontanelle were in broad agreement with a commercial foot pulse oximeter.
机译:当外周灌注较差时,需要更可靠,不具有替代测量部位,用于监测严重的新生儿中的动脉血氧饱和度。前方Fontanelle是新生儿的独特解剖学特征,已作为估计动脉血氧饱和度(SPO_2)作为替代网站。已经开发了一种新的Fontanelle PhotoPrishysmography传感器和处理系统来研究Fontanelle PhotopleSysmography(PPG)信号并在该解剖位置估计SPO_2值。初步临床试验表明,可以从新生儿Fontanelle获得具有大幅度和高信噪比的良好质量的PPG信号。估计Fontanelle的Spo_2值与商业脚脉血管计具有广泛的协议。