首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >A Clip-free Eyeglasses-based Wearable Monitoring Device for Measuring Photoplethysmograhic Signals
【24h】

A Clip-free Eyeglasses-based Wearable Monitoring Device for Measuring Photoplethysmograhic Signals

机译:基于夹层的可穿戴监控装置,用于测量PhotoPlethysmograhic信号

获取原文

摘要

An eyeglasses-based device has been developed in this work to acquire photoplethysmogram (PPG) from the nose bridge. This device is aimed to provide wearable physiological monitoring without uncomfortable clips frequently used in PPG measurement from finger and ear. Switching control is applied on the LED and photo detector for power saving. An experiment involving postural change and treadmill jogging among 10 healthy young subjects was carried out to evaluate the performance of the device. Electrocardiogram (ECG) and PPG from finger, ear and nose were simultaneously recorded, from which heart rate (HR) and pulse transit time (PTT) were calculated. The results show that PPG measured from nose and ear are more resistant to motion than signal from finger during exercise. In addition, the difference between PTT measured from ear and nose indicates that local vasomotor activities may exist on ear and/or nose channel, and suggests that PPG from different sites should be used for cuff-less PTT-based BP estimation. We conclude that this wearable device has great potential to be used in the healthcare management in the future.
机译:在这项工作中开发了一种基于眼镜的装置,以从鼻梁桥获得光学肌谱(PPG)。该装置旨在提供可穿戴的生理监测,而无需在手指和耳朵的PPG测量中经常使用的不舒服剪辑。开关控制应用于LED和照片检测器,以便省电。涉及10个健康青年对象中历史变化和跑步机慢跑的实验,以评估装置的性能。同时记录来自手指,耳和鼻的心电图(ECG)和PPG,从中计算心率(HR)和脉冲传输时间(PTT)。结果表明,从锻炼期间,从鼻子和耳朵测量的PPG比信号从手指的信号更耐受。此外,从耳朵和鼻子测量的PTT之间的差异表明耳朵和/或鼻腔通道中可能存在局部血管运动活性,并表明来自不同部位的PPG应用于基于袖口的PTT基BP估计。我们得出结论,这种可穿戴设备在未来医疗保健管理中具有很大的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号