首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >A blister-test apparatus for studies on the adhesion of materials used for neural electrodes
【24h】

A blister-test apparatus for studies on the adhesion of materials used for neural electrodes

机译:用于研究神经电极的材料粘合性的泡罩试验装置

获取原文

摘要

A blister test apparatus has been developed, which allows a quantitative adhesion analysis of thin-film metallizations on polymers manufactured in cleanroom conditions suitable for micromachining of neural electrode arrays. The device is capable of pressurizing metallic membranes at wafer level, monitoring the pressure and the height of the developing blister while detecting the moment of delamination, allowing the calculation of the adhesion energy between metal film and polymer. The machine is designed for quantitative long-term studies of materials used in neural microelectrode arrays.
机译:已经开发了泡罩试验装置,这允许在适用于神经电极阵列的微机械线的洁净室条件下制造的聚合物上的薄膜金属化的定量粘附分析。该装置能够在晶片水平下加压金属膜,监测显影泡罩的压力和高度,同时检测分层的时刻,允许计算金属膜和聚合物之间的粘合能量。该机器专为神经微电极阵列中使用的材料的定量长期研究而设计。