首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Testing of Neurosurgical Needle Steering via Duty-Cycled Spinning in Brain Tissue in Vitro
【24h】

Testing of Neurosurgical Needle Steering via Duty-Cycled Spinning in Brain Tissue in Vitro

机译:脑组织中脑组织中的神经外科针转向的测试

获取原文

摘要

A technique for steering of flexible bevel-tipped needles through tissue, providing proportional control of trajectory curvature by means of duty-cycled rotation or spinning during insertion, has been presented previously, and tested in vitro in gelatin samples. The present paper presents the results of testing under more authentic conditions. Thirty-two needle insertions were performed in cadaver brain tissue. Needle insertion was controlled interactively by the surgeon, adjusting parameters intraoperatively as needed based on fluoroscope images acquired at intervals of 1-2 cm of insertion depth. The average target acquisition error was 1.80 ± 1.33 mm.
机译:通过组织通过组织转向柔性锥形针的技术,通过在插入期间通过循环旋转或纺丝来提供对轨迹曲率的比例控制,并在明胶样品中在体外测试。本文介绍了在更真实条件下测试的结果。在尸体脑组织中进行32个针插入。通过外科医生相互作用地控制针插入,根据以1-2cm的插入深度的间隔获取的荧光镜图像,根据需要调节参数。平均目标采集误差为1.80±1.33 mm。