A Protocol Design for Evaluation of Wearable Cuff-less Blood Pressure Measuring Devices

机译:可穿戴型血压测量装置评估的协议设计

获取原文

摘要

This paper proposes one of the unique requirements in the validation protocol of the IEEE P1708 standard in assessing wearable cuff-less blood pressure (BP) measuring devices. Based on principles that are different from that of the conventional cuff-based devices, the cuff-less BP measurement approaches often require an individual calibration procedure. In this study, we used data from an experiment carried out on 28 subjects with a total of 139 sets of BP measurements as an example to show that breakdown of the performance evaluation of cuff-less devices according to the change of BP from the point of calibration is crucial for understanding and interpreting the overall accuracy of the device.
机译:本文提出了IEEE P1708标准的验证协议中的独特要求之一,用于评估可穿戴型脚踏血压(BP)测量装置。基于与传统的基于袖带的设备不同的原理,较少的BPS测量方法通常需要单独的校准程序。在这项研究中,我们使用了在28个受试者上进行的实验中的数据,共有139套BP测量作为示例,以表明根据BP的变化从较大的设备的性能评估的绩效评估分解校准对于了解和解释设备的整体精度至关重要。