首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Multi-finger structure and pulsed-powering operation scheme for CMOS LSI-based flexible stimulator for retinal prosthesis
【24h】

Multi-finger structure and pulsed-powering operation scheme for CMOS LSI-based flexible stimulator for retinal prosthesis

机译:基于CMOS LSI的柔性刺激器的多指结构和脉冲供电操作方案,用于视网膜假体

获取原文
获取外文期刊封面目录资料

摘要

Multi-finger structure was proposed to improve flexibility of the CMOS LSI-based multi-chip retinal stimulator. A dual-finger retinal stimulator was fabricated and its functionality was demonstrated in retinal stimulation experiments on rabbit's retina, we also proposed an idea of pulsed-powering operation scheme for the multi-chip flexible retinal stimulator. We compared the pulsed-powering scheme with conventional one in a simulation, and show that the pulsed-powering can be an alternative operation scheme for the neural stimulator that provides an improved safety to the biological tissue.
机译:提出了多指结构,提高基于CMOS LSI的多芯片视网膜刺激器的灵活性。制造了双指视网膜刺激器,其功能在兔视网膜视网膜刺激实验中证明,我们还提出了用于多芯片柔性视网膜刺激器的脉冲动力操作方案的思想。我们将脉冲动力方案与常规仿真进行比较,并表明脉冲功率可以是神经刺激器的替代操作方案,其为生物组织提供改进的安全性。