In this paper, an automated system for measuring ac-dc transfer differences of TVC's at frequency 1 kHz is presented. Mode of this system' functioning is based on measuring ac-dc transfer differences according to it definition. The system provides measuring ac-dc transfer differences of TVC's with uncertainty less than 10↑-6 at frequency 1 kHz.