Analysis and calculation of systemic error of voltage difference method in the calibration for heavy dc sensors

机译:重型直流传感器校准中电压差法系统误差的分析与计算

获取原文

摘要

A method of analysing and calculating the systemic error of voltage difference method in the calibration for heaby dc sensors is developed. For both current-voltage sensor and current-current sensor, the expressions for their systematic error are derived. The definition of the methokical error are discussed. The rules for the reasonable selection of parameters of the equipments used for the calibration system are given.