The objective of this research was to develop the low - cost blood pressure simulator for verifying of oscillometric automated sphygmomanometer in Thailand and compared this simulator with commercial simulator Fluke Cufflink. The testing results of the precision and the accuracy of pressure generation showed that the average percentage error and percentage precision were 0.5 % and 100 % successively. Finally, oscillometric BP devices (6 models from 5 manufacturers) were tested for checking the performance of the simulator. The results of the performance testing showed that the percentage error and precision were ol % and 99 % sequentially. We expected to continue to improve for the better quality and will be commercialized to be used to verify the automated sphygmomanometer in Thailand afterwards.
机译:这项研究的目的是开发一种低成本的血压模拟器,用于验证泰国的示波法自动血压计,并将该模拟器与商用模拟器Fluke Cufflink进行比较。精度和压力产生精度的测试结果表明,平均误差百分比和精度百分比依次为0.5%和100%。最后,对示波BP设备(来自5个制造商的6个型号)进行了测试,以检查模拟器的性能。性能测试的结果表明,百分比误差和精度依次为ol%和99%。我们预计将继续改进以提供更好的质量,并将在随后进行商业化以用于验证泰国的自动血压计。