Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator.

机译:使用血压模拟器评估A&D UA-767和Welch Allyn Spot Vital Signs无创血压监测仪。

获取原文
获取原文并翻译 | 示例
           

摘要

The performance of five units of the A&D UA-767 NIBP monitor and five units of the Welch Allyn Spot Vital Signs noninvasive blood pressure monitor was evaluated with the Biotek BP Pump blood pressure simulator under a variety of conditions. Using the simulator to provide a normal blood pressure waveform at 80 bpm over a range of pressures, it was found that the mean bias for the combined results from the A&D monitors was 1.9+/-2.8 mmHg and from the Welch Allyn monitors was 0.7+/-2.4 mmHg. No individual measurement showed a bias greater than 10 mmHg. A bias of greater than 5 mmHg was present in 28 out of 150 measurements for the A&D monitor and 10 out of 150 measurements for the Welch Allyn monitor. These results are comparable with ratings achieved by the instruments when tested previously according to the British Hypertension Society protocol, but testing with a simulator allowed assessment of aspects of performance which were not included in the British Hypertension Society protocol.
机译:使用Biotek BP Pump血压模拟器在各种条件下评估了五台A&D UA-767 NIBP监护仪和五台Welch Allyn Spot Vital Signs无创血压监护仪的性能。使用模拟器在一系列压力范围内以80 bpm的速度提供正常血压波形时,发现A&D监护仪和Welch Allyn监护仪的综合结果的平均偏差为1.9 +/- 2.8 mmHg。 /-2.4毫米汞柱。没有单独的测量显示偏差大于10 mmHg。 A&D监视器的150个测量结果中有28个偏差大于5 mmHg,Welch Allyn监视器的150个测量结果中有10个偏差大。这些结果与以前根据英国高血压学会的协议进行测试时仪器所获得的评级相当,但是使用模拟器进行测试可以评估英国高血压学会协议中未包括的性能方面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号