首页> 外文会议>Computing in Cardiology Conference >Multicenter Validation of Three-Dimensional Echocardiographic Quantification of the Left Heart Chambers using Automated Adaptive Analytics
【24h】

Multicenter Validation of Three-Dimensional Echocardiographic Quantification of the Left Heart Chambers using Automated Adaptive Analytics

机译:使用自动适应性分析的左心室三维超声心动图定量的多中心验证

获取原文

摘要

Although recommended by the current guidelines, 3D echocardiographic (3DE) quantification of the cardiac chambers in clinical practice has been lagging, because of time-consuming analysis. We recently validated an automated algorithm that measures left atrial (LA) and left ventricular (LV) volumes and ejection fraction (EF). This study aimed to determine its accuracy and reproducibility in a multicenter setting. 180 patients underwent 3DE imaging (Philips) at 6 sites. Images were analyzed using automated HeartModel (HM) software with endocardial border correction when necessary, and also by conventional manual tracing. Measurements were performed independently by each site and by the Core Laboratory (CL), whose measurements were used as a reference. Inter-technique comparisons included HM measurements by the sites with and w/o corrections against manual tracing by CL. Intra-technique comparisons included HM measurements by the sites against those by CL (with and w/o corrections, respectively). Inter-technique comparison w/o corrections showed strong correlations, with the automated technique slightly underestimating LV volumes. Corrections were either unnecessary or minimal in most patients, and affected the measurements minimally. Intra-technique comparisons with corrections showed better correlations and smaller biases than the inter-technique comparison. All automated measurements with corrections were more reproducible than manual measurements. Automated volumetric analysis of left-heart chambers is an accurate alternative to conventional methodology, which yields almost the same values across laboratories and is more reproducible. This technique may contribute towards full integration of 3DE quantification into clinical routine.
机译:虽然目前的准则推荐,但由于耗时的分析,临床实践中心室的3D超声心动图(3DE)量化一直滞后。我们最近验证了一种自动算法,可测量左心房(LA)和左心室(LV)体积和喷射分数(EF)。本研究旨在确定多中心设置中的准确性和再现性。 180名患者在6个地点进行3DE成像(飞利浦)。在必要时使用自动心脏模型(HM)软件进行分析图像,并且在必要时,以及传统的手动跟踪。通过每个位点和核心实验室(CL)独立进行测量,其测量用作参考。技术间比较包括与CL的手动跟踪的网站和W / O校正的HM测量。技术内部比较包括由CL(分别为和W / O校正)对这些位点的HM测量。技术间比较W / O校正表现出强烈的相关性,自动化技术略微低估了LV体积。在大多数患者中,校正是不必要的或最小的,并影响测量。与校正的技术帧内比较显示出比技术间比较更好的相关性和较小的偏差。所有具有校正的自动测量比手动测量更可重复。左心室的自动体积分析是对传统方法的准确替代方案,其在实验室中产生几乎相同的值,并且更可重复。该技术可能有助于将3DE定量完全集成到临床常规中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号