Corrosion Response of Cobalt Chrome - Titanium 6Aluminum 4Vanadium Modular Taper Connections in Revision Constrained Condylar Knee Femoral Constructs

机译:钴铬的腐蚀响应 - 钛6 alumumum 4Vanium锥形锥形锥形锥形锥形连接约束约束髁臂股骨构建体

获取原文

摘要

Based on the review by Amhost et. al. [2] of 144 modular clinical retrievals across a range of manufacturers, using the Modified Goldberg Visual Fretting Corrosion Scale, evidence of taper damage is not uncommon in modular total knee arthroplasty interfaces. The corrosion scores obtained from the evaluation of the subject CoCr - Ti6A14V revision CCK construct appear consistent with those of commercially available modular femoral implants. Modular connections using CoCr-Ti6A14V did not exhibit significantly different visual corrosion scores from those utilizing only Ti6A14V, suggesting the impact of design, in addition to material selection, on the corrosion rate.
机译:根据Amhost等的审查。 al。 [2] 144个模块化临床检索在一系列制造商中,使用改良的Goldberg视觉烦恼腐蚀量表,锥度损坏的证据在模块化全膝关节形成术界面中并不少见。从受试者COCR-TI6A14V修正CCK构建体获得的腐蚀分数看起来与市售的模块化股植入物的评估结果一致。使用COCR-TI6A14V的模块化连接没有从仅使用TI6A14V的那些具有显着不同的视觉腐蚀分数,这表明设计的影响,除了材料选择,腐蚀速率。