Inter-pixel sensitivity variations in the X-ray flat panel detector give rise to gain non-uniformity. These artifacts degrade the quality of digital radiography and fluoroscopy imaging. The state-of-the-art gain correction methods are analyzed and found to be inadequate in dealing with beam hardening-related non-uniformity. We have developed a multi-dimensional signal model to accommodate the beam hardening related factors in practical scenarios. A novel beam hardening-respecting (BHR) gain correction method is devised for medical X-ray flat panel detectors. Improvements of the proposed technique over fixed point and variable gain correction are demonstrated.
机译:像素间敏感性变化在X射线平板检测器中产生不均匀性。 这些伪像降低了数字射线照相和荧光检查成像的质量。 分析最先进的增益校正方法,发现在处理与光束硬化相关的不均匀性的情况下是不充分的。 我们开发了一种多维信号模型,以在实际情况下容纳光束硬化相关因素。 为医用X射线平板探测器设计了一种新型光束硬化 - 方向(BHR)增益校正方法。 证明了固定点和可变增益校正的所提出的技术的改进。