In this paper, a piezoelectric micromachined ultrasound transducer (PMUT) array for multi-frequency imaging is investigated. The PMUTs are thoroughly characterized and a finite element model is created in COMSOL Multiphysics® software. First, an axisymmetric model of a single PMUT is constructed that couples the solid mechanics, electrostatics and pressure acoustics interfaces. The model accurately predicts static deformation, resonance frequencies and displacement at resonance. Finally, a full 3D model is created to investigate the origin of an observed resonance mode in the array that does not correspond to any flexural mode of the PMUTs. Acoustic crosstalk through the fluid is identified as the likely cause for this “array mode”. Acoustic coupling should therefore be considered carefully in future designs and utilized where possible to achieve improved performance or added functionality such as multi-frequency imaging.
机译:本文研究了用于多频成像的压电微机械超声换能器(PMUT)阵列。 PMUT彻底表征,并在COMSOL Multiphysics中创建有限元模型® 软件。首先,构造单个PAUT的轴对称模型,其耦合固体力学,静电和压力声学界面。该模型精确地预测静态变形,谐振频率和谐振的位移。最后,创建完整的3D模型以研究不对应于PMUT的任何弯曲模式的阵列中观察到的谐振模式的原点。通过流体的声学串扰被识别为该“阵列模式”的可能原因。因此,应在未来的设计中仔细考虑声学耦合,并在可能实现改进的性能或添加功能,例如多频成像。