An Ultra-High Element Density pMUT Array with Low Crosstalk for 3-D Medical Imaging

机译:具有低串扰的超高元素密度pMUT阵列,用于3-D医学成像

获取原文
   

获取外文期刊封面封底 >>

       

摘要

A ∼1 MHz piezoelectric micromachined ultrasonic transducer (pMUT) array with ultra-high element density and low crosstalk is proposed for the first time. This novel pMUT array is based on a nano-layer spin-coating lead zirconium titanium film technique and can be fabricated with high element density using a relatively simple process. Accordingly, key fabrication processes such as thick piezoelectric film deposition, low-stress Si-SOI bonding and bulk silicon removal have been successfully developed. The novel fine-pitch 6 × 6 pMUT arrays can all work at the desired frequency (∼1 MHz) with good uniformity, high performance and potential IC integration compatibility. The minimum interspace is ∼20 μm, the smallest that has ever been achieved to the best of our knowledge. These arrays can be potentially used to steer ultrasound beams and implement high quality 3-D medical imaging applications.
机译:首次提出了具有超高元素密度和低串扰的〜1 MHz压电微机械超声换能器(pMUT)阵列。这种新颖的pMUT阵列基于纳米层旋涂铅锆钛钛膜技术,可以使用相对简单的工艺以高元素密度制造。因此,已经成功地开发了关键的制造工艺,例如厚压电膜沉积,低应力Si-SOI键合和体硅去除。新型的细间距6×6 pMUT阵列都可以以所需的频率(〜1 MHz)工作,具有良好的均匀性,高性能和潜在的IC集成兼容性。最小的间距约为20μm,这是据我们所知所达到的最小间距。这些阵列可潜在地用于操纵超声波束并实现高质量的3D医学成像应用。