During this study the distortion phenomenon was investigated on selected UHMWPE cups removed from patients, using 3dimensional measurement method. The cups investigated were machined by Metrimed Ltd. From ram extruded Chirulen 1020 type UHMWPE rods. The samples were measured with a 3D touch coordinate measurement device with an accuracy of 3 microns. It has been found that in an average use of 10 years the geometric change of the cups is about 0,5mm, and the distortion phenomenon of cups in males v.s. females were found to be different. The failure processes of the samples retrieved from males and females show different wear/creep behavior depending on gender, implicating that the differences in movement dynamics and anatomy have different effects on the distortion phenomena of the cups. Further studies are necessary on more specimens in order to give more detailed analysis.
机译:在此期间,使用3DImensional测量方法研究了从患者中取出的选定的UHMWPE杯中研究了失真现象。研究的杯子由Metrimed Ltd.加工来自Ram挤压的胆管1020型UHMWPE棒。用精度为3微米的3D触摸坐标测量装置测量样品。已经发现,在平均使用10年的情况下,杯子的几何变化约为0.5mm,以及男性V.S的杯子的变形现象。发现女性不同。从雄性和女性中检索的样品的失败过程显示了根据性别的不同磨损/蠕变行为,这意味着运动动态和解剖学的差异对杯子的失真现象具有不同的影响。进一步的研究是对更多标本进行进一步的研究,以便提供更详细的分析。