首页> 外文会议>Hungarian conference on materials science >In vivo failure phenomena of ultra-high molecular weight polyethylene acetabular cups
【24h】

In vivo failure phenomena of ultra-high molecular weight polyethylene acetabular cups

机译:超高分子量聚乙烯髋臼杯的体内破坏现象

获取原文

摘要

During this study the distortion phenomenon was investigated on selected UHMWPE cups removed from patients, using 3dimensional measurement method. The cups investigated were machined by Metrimed Ltd. From ram extruded Chirulen 1020 type UHMWPE rods. The samples were measured with a 3D touch coordinate measurement device with an accuracy of 3 microns. It has been found that in an average use of 10 years the geometric change of the cups is about 0,5mm, and the distortion phenomenon of cups in males v.s. females were found to be different. The failure processes of the samples retrieved from males and females show different wear/creep behavior depending on gender, implicating that the differences in movement dynamics and anatomy have different effects on the distortion phenomena of the cups. Further studies are necessary on more specimens in order to give more detailed analysis.
机译:在这项研究中,使用3维测量方法对从患者身上取出的选定UHMWPE杯进行了变形现象的研究。所研究的杯子是由Metrimed Ltd.从拉挤的Chirulen 1020型UHMWPE棒材加工而成的。使用3D触摸坐标测量设备以3微米的精度测量样品。已经发现,在平均使用十年中,杯子的几何变化约为0.5mm,并且在男性对女性中杯子的变形现象。发现女性不同。从性别中提取的样品的失效过程显示出不同的磨损/蠕变行为,这取决于性别,这表明运动动力学和解剖结构的差异对杯子的变形现象有不同的影响。为了提供更详细的分析,有必要对更多的标本进行进一步的研究。