【24h】

Monte Cario Simulation of a Micro-multileaf Collimator

机译:微型多叶准直器的蒙特卡罗模拟

获取原文

摘要

The performance of a dynamic micro-multileaf collimator manufactured by 3D-Line, for stereotactic radiosurgery has been studied. Monte Cario simulations based on PENELOPE code and measurements with three different detectors (PTW Semiflex 31010 chamber, PTW PinPoint 31016 chamber and PTW Diode 60008) have been carried out. The geometry of the micro-multileaf collimator was tested by calculating the transmission through it and it was needed to reduce the separation between the leafs indicated by the manufacturer to reproduce the experimental results. An approximate simulation in which only Compton interactions of photons are considered was analyzed, providing a good agreement with the full simulations. With the Monte Cario model fixed, output factors for various field sizes were calculated and compared to the experimental ones, obtaining a good agreement. Percentage depth doses and transverse profiles at two depths measured with the Diode for small fields were well reproduced by the simulation, while the measurements performed with the PinPoint chamber showed differences in the Percentage depth doses, at large depths, and in the transverse profiles, at the penumbra. Monte Cario simulations and Semiflex and Diode measurements, performed for a 7 cm × 7 cm field, were in nice agreement, while those obtained with the PinPoint chamber showed differences that increased with the depth in water. At the phantom entrance, all measurements showed non-negligible differences what made Monte Cario a good option to estimate the absorbed dose in this region.
机译:研究了由3D-Line制造的动态微型多叶准直仪用于立体定向放射外科手术的性能。进行了基于PENELOPE代码的蒙特卡里奥模拟,并使用三个不同的检测器(PTW Semiflex 31010腔室,PTW PinPoint 31016腔室和PTW Diode 60008)进行了测量。通过计算微多准直器的透射率来测试其几何形状,并且有必要减少制造商指示的叶片之间的间隔,以重现实验结果。分析了仅考虑光子的康普顿相互作用的近似模拟,与完整模拟提供了良好的一致性。在固定蒙特卡里奥模型的情况下,可以计算出各种场大小的输出因子,并将其与实验值进行比较,从而获得了很好的一致性。通过模拟可以很好地再现用二极管针对小场测量的两个深度处的深度剂量百分比和横向轮廓,而使用PinPoint腔室进行的测量表明,在大深度处以及横向轮廓处,深度处的百分比深度剂量存在差异半影。在7 cm×7 cm的电场下进行的蒙特卡里奥模拟以及Semiflex和Diode的测量结果非常吻合,而使用PinPoint腔室获得的结果显示差异随水深的增加而增加。在幻影入口处,所有测量结果均显示出不可忽略的差异,这使得Monte Cario成为估算该区域吸收剂量的理想选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号