Evaluation of Hemolysis Caused by a Miniature Heart Catheter Pump

机译:用微型心脏导管泵引起的溶血评估

获取原文

摘要

An important performance criterion for blood pumps is blood damage due to the non-physiological loads on the blood cells. Especially for miniature pumps that are inserted into blood vessels in catheter devices and - due to their small diameters - operate at high rotational speeds, hemolysis is an important design issue. The Assistocor heart catheter pump is a left ventricular assist device which is inserted into the Aorta and has an outer diameter of 6.8 mm, it operates at rotational speeds up to 40000 rpm. In-vitro tests with animal blood and different pump prototypes of the Assistocor heart pump showed that hemolysis values stayed below the recommended values for five out of nine of the tested prototypes and are comparable with results from literature, indicating that the design process is adequate to keep the damage of blood cells within acceptable limits despite the high rotational speeds.
机译:由于血细胞上的非生理载荷,血液泵的重要性能标准是血液损伤。特别是对于插入导管器件中的血管中的微型泵和 - 由于它们的小直径 - 在高转速下运行,溶血是重要的设计问题。辅助孔心脏导管泵是左心室辅助装置,该辅助装置插入主动脉,外径为6.8mm,它以高达40000rpm的旋转速度运行。与动物血液和不同泵原型的体外测试助源泵心脏泵显示,溶血值低于测试的原型中五个中的五个出来的推荐值,并且与文献的结果相当,表明设计过程足够了尽管旋转速度高,但在可接受的限度内保持血细胞的损伤。