首页> 外文会议>International Conference on Advances in Electrical Engineering >Effect of pump characteristic (H-Q) curves on the efficiency of Total Artificial Heart
【24h】

Effect of pump characteristic (H-Q) curves on the efficiency of Total Artificial Heart

机译:泵特性(H-Q)曲线对总人造心脏效率的影响

获取原文

摘要

Nowadays, heart failure of human being is a very common occurrence all over the world. Total Artificial Heart (TAH) is an outstanding achievement of biomedical engineering where the patient's natural heart is removed and replaced with two mechanical pumps that provide perfusion to the systemic and pulmonary circulations. This paper focuses on the effect of pump characteristic (H-Q) curves on the efficiency of TAH. To show the effect of pump parameters, a simplified model of human cardiovascular system (CVS) is developed with an electrical circuit. A numerical simulation is conducted in order to find out the different physiological pressures and pump flows in response to change of parameters related to pump performance curves. Finally, the effect of H-Q curves on the efficiency of TAH is analyzed.
机译:如今,人类的心力衰竭是世界各地的常见发生。人工心脏(TAH)是对生物医学工程的突出成就,其中除去患者的天然心脏并用两个机械泵代替,为全身和肺循环提供灌注。本文侧重于泵特性(H-Q)曲线对TAH效率的影响。为了显示泵参数的效果,用电路开发了一种人体心血管系统(CVS)的简化模型。进行数值模拟,以便响应于与泵性能曲线相关的参数的变化,找到不同的生理压力和泵流。最后,分析了H-Q曲线对TAH效率的影响。